The CPFM-8800 series is a family of advanced flow meters based on the Coriolis principle fabricated exclusively from PFA (Perfluoroalkoxy) polymeric material. CPFM-8800 flow meters are comprised of two assemblies: one containing the sensor, the other containing the supporting electronics. CPFM-8800 sensors are specially designed for measuring liquids in high-purity semiconductor, bio-pharmaceutical and other applications that require all PFA-wetted surfaces and provide a Mass Flow Rate, Total Mass and Temperature with accuracy ±1% of reading .
This product is protected by U.S. and International patents. See Malema's list of patents for more information.
Designed and Assembled in California.
Mass Flow Ranges
Model 8803-1: 50-1500 g/min*
Model 8803-2: 150-4000 g/min*
Zero Offset Stability (Z.O.S.)
Model 8803-1: 2 g/min
Model 8803-2: 3 g/min
Accuracy
± {1% of rate + Z.O.S.}
Temperature
Ambient: 0°–50°C; Fluid 18°–50°C
Operating Pressure
80 PSIG (Max.)
*Pressure drop at Max. Flow range = 10 psi on water (1cP)
Supply Voltage
24VDC ±10%
Power Consumption
Max 6W
Programming
Operator Parameter configuration through USB interface with a PC
Output Interfaces
4-20 mA Current Loop, Digital I/O
LCD Display
2 lines; 16 characters per line
Analog Output Module
4-20mA; 500 Ohms max load
Digital Input/Output Module
Configurable as Frequency or Digital I/O
Frequency Output
0 to 10KHz proportional to flow rate
Process Connections
3/8” tube connection, 1/4" tube connection **
Wetted Material
PFA
Sensor Dimensions
L(254mm) x W(73mm) x H(115mm)
Transmitter Dimensions
L(210mm) x W(58.5mm) x H(133.5mm)
Weight
Sensor: 0.86 Kg; Transmitter: 1.22 Kg
Cable Length
Up to 30m max cable length between sensor and electronics assemblies
** Consult factory for other types of process connection requirements.
For additional specification details, please refer to the product data sheet located in the documentation tab of this web page.